Search results for " CAD modeling"

showing 1 items of 1 documents

Virtual prototyping of a new intramedullary nail for tibial fractures

2012

In this work a new distal interlocking system has been developed which is easy to use, allows a reduction of the operating time and consequently the exposure to radiations both for surgeon and patient. The main goal of this study has been the design of a new intramedullary nail for tibial fractures able to simplify and speed up the distal locking operation phases. After a preliminary stage during which several candidate concepts have been proposed and analysed, the best solution has been developed and deeply investigated. The new system, called “expansion nail”, has been firstly modelled by setting up a full parametric CAD model and, then, tested by running non linear FEM analyses to evalua…

Engineering drawingEngineeringbusiness.industrymedicine.medical_treatmentWork (physics)Structural engineeringIndustrial and Manufacturing Engineeringlaw.inventionIntramedullary rodlawMechanical stabilityModeling and SimulationOperating timeNail (fastener)medicineVirtual Reality tibial fractures CAD modeling FEMbusinessSettore ING-IND/15 - Disegno E Metodi Dell'Ingegneria IndustrialeReduction (orthopedic surgery)InterlockingVirtual prototyping